Anasayfa / Ürünlerimiz / 1090 Portable Process Calibrator
Product Description:
The 1090 is a precision portable process calibrator that combines source and measurement functions for thermocouples, PT100s, µV, mV & mA.
As a multifunction instrument the 1090 combines accuracy and durability with simple operation, making it ideal for process plant applications.
Thermocouple Measurement and Simulation
Measure and simulate the temperature and mV characteristics of J, K, T, R, S, B, N and E thermocouples.
Cold Junction Compensation
The unit can be operated with or without internal cold junction compensation.
PT100 Measurement and Simulation
Based on 0.3850 alpha probe standard. Range is -200°C to 700°C.
Measurement and Source (µV, mV, and mA)
Measurement ranges are 0 to +/-30mV and 0 to +/-60mA. Source ranges are 0 to +/-80mV and 0 to +80mA.
Temperature Units Selection
The display can be easily changed from °C to °F. The equivalent uV (thermocouples) and ohms (PT100) can also be shown.
24V Process Loop Drive Mode
A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA (or a lower level if required).
Inching (Incrementing/Decrementing)
The unit has a general-purpose inching function. This adjusts the output in fixed increments of temperature (thermocouples only) or voltage or current. The set-up menu gives a the user a choice of three levels of increment i.e. 0.1, 1 or 10 for °C/°F, or 1, 10, or 100 uV/uA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. This is especially useful for calibrating thermostat controllers that have tight specification on hysteresis.
Memory Recall and Step/Auto-Step Functions
Up to 10 values can be stored in the unit’s non-volatile memory and they can be recalled at any time. The user can also manually step through them in sequence using the step key. Continuous stepping (auto-step) is also available at any user selectable rate between 1 & 10 seconds/step.
Power is via an internal high capacity rechargeable metal hydride battery that can be recharged from the supplied external mains charger. An auto power-down feature helps conserve battery life by switching off the instrument if inactive for over 5 minutes. This feature can be disabled if not required. The unit is supplied in a robust case with a carrying strap. A pocket for the instruction manual is provided.
Product Specification:
TEMPERATURE MEASURE
THERMOCOUPLE TYPE TEMPERATURE RANGE (°C) ACCURACY (°C)
J -200 to 580 0.7
K -200 to -150
-150 to 750 2.5
0.5
T -200 to 0
0 to 400 1.5
0.4
R 50 to 400
400 to 1750 3.0
1.5
S -50 to 100
100 to 1750 3.0
1.5
B 110 to 1000
1000 to 1800 3.5
1.5
N -100 to 890 0.6
E -50 to 400 0.4
Resolution 0.1°C or 0.1°F
TEMPERATURE SİMULATİON
THERMOCOUPLE TYPE TEMPERATURE RANGE (°C) ACCURACY (°C)
J -210 to 150
150 to 1200 0.15
0.3
K -270 to 190
190 to 1250 0.5
0.4
T -200 to 150
150 to 400 0.4
0.5
R -50 to 800
800 to 1750 0.8
2.0
S -50 to 850
850 to 1750 0.9
2.0
B 100 to 1200
1200 to 1800 2.0
3.0
N -270 to 260
260 to 1300 0.5
1.0
E -50 to 1000 0.3
Resolution 0.1°C or 0.1°F
Note: An additional correction representing the equivalent 1μV should be allowed for stray thermal emf effects (Temperature measure & simulation).
PT100 SİMULATİON
14 SET TEMPERATURE POİNTS (°C) ACCURACY
-100, -50, -20, 0, 20, 50, 100, 200, 300, 400, 500, 600, 700, 800 0.1% of resistance value (typically 0.5 °C)
PT100 MEASURE
RANGE ACCURACY RESOLUTİON
200 to 700 °C, 2 wire. 0.2% of resistance value (typically 0.7 °C) 0.2°C (or 0.2°F)
MİLLİVOLT MEASURE
RANGE RESOLUTİON ACCURACY INPUT RESİSTANCE
0 to +/- 30mV 10μV 0.05% of full scale ±1 digit 100kΩ
MİLLİVOLT SOURCE
RANGE RESOLUTİON ACCURACY OUTPUT RESİSTANCE
0 to 8mV 0.5μV +/- 4μV 10Ω
8 to 80mV 5μV 0.02% of full scale 10Ω
MİLLİAMP MEASURE
RANGE RESOLUTİON ACCURACY INPUT RESİSTANCE
0 to 60mA 20μA 0.05% of full scale ±1 digit 0.5Ω
MİLLİAMP SOURCE
RANGE RESOLUTİON ACCURACY
0 to 8mA 0.5μA +/-10μA
8 to 80mA 5μA 0.02% of full scale
FURTHER FEATURES AND FUNCTİONS
Inching Three levels of increment, 0.1, 1 or 10 for °C/°F, and 1, 10, or 100 μV/ μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output.
24V Process Loop Drive A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA
(or a lower level if required). Max Load Specifications: 300Ω/80mA, 480Ω/50mA, 1.2kΩ/20mA
Memory recall & step 10 memory locations for non-volatile storage of values. Manual & AutoStep, rate adjustable from 1 to 10 sec/step
Cold Junction Compensation Accuracy 0.2°C. Resolution 0.1°C.
Operating Temperature -10 to 40°C (15 to 105°F)
Connections Industry standard 4mm screw terminals
Power A Ni-MH rechargeable battery pack gives approx. 60 hours continuous use. The mains recharger supplied gives full recharge in 11 hours. To conserve battery life a user inactivity power-down feature is included.
Dimensions H235 x W150 x D75mm
Weight 1.2kg
Supplied With Carry case, rechargeable battery pack (installed) with mains charger, and user manual
Optional Extras Calibration certificates traceable NPL and UKAS
Country of Origin UK